AC Hysteresis Graph Plotter

Location:
New , Delhi India
Status:
New
Sample Availibility:
yes
Payment Terms:
Cash, Bank Transfer, Letter of Credit, T/T, L/C

Product Description

Under continuous frequency of 1 kHz to 500 kHz, automatic measurement of magnetic hysteresis loop of the soft magnetic materials such as silicon steel, precision alloy, non-crystal and ferrite, accurate measurement on the dynamic magnetic characteristic parameters such as amplitude permeability, loss angle loss iron Pc, remanence Br and coercivity Hc Windows measurement software applied simply. It conforms to National Standards GB 5026-85, GB 3658-90, GB/T 9632.1-2002, GB/T 19346-2003, industry standard SJ/T 10281-91 and international standard IEC 60367-1, IEC 60404-6.The controlling of computer and A/D sampling replace the traditional analog bridge, frequency meter, ammeter, voltage meter, power meter, the whole testing process is automatic completed.

General Features

Testing sample varieties: soft magnetic ferrite, permalloy, amorphous and nano crystal.Test sample shapes: toroid, E and U shapes.Test samples with closed magnetic circuit can be directly winded and measured on samples. Magnetizing coil (N1) and measuring coil (N2) form a no-load transformer.Non-inductive resistance connected on magnetizing coil loop to determine magnetizing current and magnetic field intensity through the measurement of pressure drop on non inductive resistance, magnetic field peak value can be locked through digital feedback, magnetic field lock precision 0.5%.Magnetic induction obtained through digital integration of coil measuring voltage, magnetic induction peak value locked through digital feedback, magnetic induction lock precision 0.5%.Power source and sampling amplifier integrated into a casing with simple interface: one RS232 interface connected to the computer, two way voltage alarm connected to high speed A/D card.Different power sources can be selected according to different test requirements: 1 kHz ~ 300 kHz or 500 kHz, 5 kHz or 10 kHz ~ 100 kHz or other frequency ranges customized.Measuring dynamic hysteresis loop through volammetry and digital integration can accurately measure dynamic magnetic characteristic parameters such as µa, d, Pc, Br and Hc, and calculate µ', µ'', µL, µR, Q and AL, etc. Automatic and continuous measurement up to 255 testing points, the testing time of every testing point is about 12 seconds, fixed-frequency or fixed Bm optional in multipoint test.Powerful function of the software, the technical request to the testing staff is very low.Operate under Windows 95/98, Windows 2000/XP operating systems, conform to Windows software specification, visual and simple operation.Full automatic control, intelligent identification, users' interventions are unnecessary in the whole testing process.Sampling waveform and instrument status under real-time monitoring, and can be stopped at any time.The effective sectional area and length of magnetic path of samples can be calculated according to the overall dimension of samples.File system adopts database format, can directly print or output test result to Excel form.Powerful file management functions: data saving and deleting, remove all the data etc.Data files contain complete sampled data, sample parameters, instrument parameters and testing schemes, adopt text format, can be conveniently typed into other software.Display I(t), U(t) and B(t) sampling waveform and B(H) hysteresis loop, and coordinate message at every data point on the curve.Multipoint cluster, can display B(H) magnetic hysteresis cycle cluster, B(H) magnetization curve, Ã,¼a(H) permeability curve, ua(F) permeability curve, Ps(B) damage curve and Ps(F) damage curve, and can display coordinate message at every data point on the curve.B(H) magnetization curve, ua(F) permeability curve, Ps(B) damage curve and Ps(F) damage curve measured under different conditions can be combined on one drawing for comparative analysis.Set up upper and lower limits according to ua, Ps, Bm, Br, Hc and Hm, etc. Qualify test result and determine through the color of data table.Support various models of printers, test report accurately match printer sheet.Print preview function, it can conveniently regulate the size and edge distance of test report.Test report can be directly printed, or generate JPG image file.JPG image file can be sent directly through E-mail, or saved into disk.Test reports contain complete curve diagrams, test results, test conditions and sample parameters. Both Chinese and English format can be adopted to facilitate adding user mark and enterprise nameMultiple unit systems optional to satisfy the application habits of different users.


System Specifications

Adopt volammetry to measure loss Ps and amplitude permeability µa and loss angle , when test frequency is 50 kHz, conduct test on ferrite circular ring sample with standard size (d1 = 40 mm, d2 = 32 mm, h = 6 ~ 10 mm), technical indices as follows:

Parameters measured

Bm (%)

Br (%)

Hc (%)

ua (%)

Ps (%)

(%)

Uncertainty (k = 2)

1

1.5

2

3

5

5

Repeatability (constant temperature)

0.5

1

1

2

3

2

Instrument Specifications

TPS-200SA/300k Soft Magnet AC Test Power SupplyTPS-200SA/500k Soft Magnett AC Power Supply
Output Power: 200 VA sine waveOutput Power: 200 VA sine wave (below 300 kHz)
Frequency Range: 1 kHz ~ 500 kHz Frequency Range: 1 kHz ~ 500 kHz
Frequency Fineness: < 0.1% * Current ValueFrequency Fineness: < 0.1% * Current Value
Frequency Error: < 0.05%Frequency Error: < 0.05%
Output Voltage: 0 ~ 50 V (Peak Value)Output Voltage: 0 ~ 50 V (Peak Value)
Voltage Fineness: Program Control 1 mV, panel < 10 mVVoltage Fineness: Program Control 1 mV, panel < 10 mV
Voltage Distortion Factor: Superior to 1%Voltage Distortion Factor: Superior to 1%
Voltage Stability: Superior to 0.1% Voltage Stability: Superior to 0.1%
Sampling Current: 0 ~ 5 A (Peak Value)Sampling Current: 0 ~ 5 A (Peak Value)
Sample Voltage: 0 ~ 10 V (Peak Value)Sample Voltage: 0 ~ 10 V (Peak Value)
TPS-500A Broadband Programmable Power supply PC5012 A/D Card
Output Power: 500 VA sine waveSampling Rate: 50 MHz * 2 Channels
Frequency Range: 10 kHz ~ 100 kHzResolution and Linearity: 12 Bits ±1/2 LSB
Frequency Fineness: < 0.1% * Current ValueVoltage Range: ±0.1 V ~ ±20 V
Frequency Error: < 0.05%Sampling Clock: 25 ns ~ 5 ms Hardware Clock
Output Voltage: 0 ~ 50 V~ 100 V~ 150 V~ 300 V Internal Storage Capacity: 4 M Bytes * 2
Voltage Fineness: Program Control 1 mV, panel < 10 mVBus Structure: PCI busbar

Voltage Distortion Factor: Superior to 1%
Voltage Stability: Superior to 0.1%
Sampling Current: 0 ~ 10 A (Peak Value)
Sample Voltage: 0 ~ 200 V (Peak Value)

Basic Configuration

Serial No.Product Name ModelQuantity
1Test Power Supply TPS-200SA 1
2Data Acquisition Card PC5012 1
3 Measurement Software SMTest 1
4Computer Dell Dimension E-520n 1
5Printer HP1020 laser printer 1
6Standard Specimen Ferrite 1


This member's information has NOT been authenticated or verified by importexportleads.net or any third party. Only Gold Members have completed an authentication and verification procedure conducted by third-party credit agencies.

For more information on importexportleads Gold Membership, Click here

Contact Information

Company:
 ACMAS TECHNOCRACY PRIVATE LIMITED
Phone Number:
  91-11-47619688
Fax Number:
 91-11-47619788